Products

Local calibration

To obtain JCSS certification for the first local thermometer calibration in Japan
In 2006, the Yamari Industries Standards Office was certified for the first local calibration in Japan under the “Temperature/Contact Thermometer” category of the calibration operator registration system (JCSS) stipulated in the Measurement Law. This has established traceability for local calibration of platinum resistance thermometers, thermocouples, and thermometers with indicator in the temperature range from-30℃ to 1100℃.
There are many requirements for on-site calibration to calibrate thermometers in use on-site. Benefits include: “The use of thermometers can be shortened, and the wiring (or indicator) can not be removed at an operation site. Calibration of thermometers, lead wires, and measuring instruments in a loop state enables the cost and labor to be reduced and the uncertainty of the measured temperature to be clarified” by calibrating thermometers and measuring instruments individually.

 

Certification range:-30℃ to 1100℃
It can be used for certified calibrations according to the demands of the site.

Calibration target thermometer*1 Specifications Type of calibration Calibration temperature(℃) Maximum capacity(℃) Reference standard
Thermocouple R, S, B, N, K, E, J, T Comparative calibration -30~250 0.5 Platinum resistance thermometer
250~550 0.7
200~1100 1.4 R thermocouple
Platinum resistance thermometer 4-wire Comparative calibration
[Resistance]
-30~140 0.08 Platinum resistance thermometer
140~250 0.4
250~550 0.6
3-wire Comparative calibration
[Resistance]
-30~140 0.08
140~250 0.4
250~550 0.6
Thermometer with indicator Minimum scale 0.1°C or more Comparative calibration -30~140 0.2 Platinum resistance thermometer
140 ~250 0.4
250~550 0.6
200~1100 1.4 R thermocouple

*1 There are restrictions on the shape of the temperature sensor.
Sensor dimensions: Outer diameter 8mm or less Length 300mm or more
Power supply: Max. 5 rated 100VAC 10A must be provided.
Ambient environment: This product cannot be used in the following places.
○ Wet in water, high moisture
○ The temperature is high. Receive direct sunlight
○ Experienced wind is flowing
○ Water-soluble or corrosive gas or dust
○ Severe vibration and noise

Contact us for calibration fees.

【英語版】山里産業コーポレートサイト
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.