Knowledge
Dimensions of the object that can be calibrated
Calibrated thermometer | Type of calibration | Calibration temperature(℃) | Dimensions of the target thermometer | Equipment used | |
---|---|---|---|---|---|
Outer diameter (mm) | Length(mm) | ||||
Thermocouple Platinum resistance thermometer Thermometer with indicator |
Comparative calibration | -40,0,100,200 | Sheath 8.0 or less Element wire 3.2 or less |
300 or more | Liquid bath |
-196~-80 | 230 or more 750 or less Or 850 or more |
Drivelock tank | |||
-100~250 (-100~-80) |
Sheath 8.0 or less Element wire 3.2 or less (Sheath 7.0 or less) |
350 or more | Liquid bath and Drivelock tank |
||
250~660(arbitrary temperature) | Sheath 8.0 or less Element wire 3.2 or less |
400 or more | Powder tank | ||
100~1100(arbitrary temperature) | 350 or more | Tubular furnace | |||
1100~1372(arbitrary temperature) | 600 or more | ||||
1100~1554(arbitrary temperature) | Element wire0.5 or less | 550 or more |
※ Shape… The length portion of the dimension should have a straight extension shape.
※ Terminal: Have a (lead) terminal that can be engaged and connected with the alligator.
※ Terminal accessories: Accessories such as terminal box and connector can be removed.
Please contact us for thermometers of dimensions and shapes other than those listed above.
Calibrated thermometer | Type of calibration | Calibration temperature(℃) | Dimensions of the target thermometer | Equipment used | |
---|---|---|---|---|---|
Outer diameter (mm) |
Length (mm) |
||||
Thermocouple Platinum resistance thermometer Thermometer with indicator |
Comparative calibration | 300 ~ 1100(arbitrary temperature) | 7.5 or less | 200 or more | Spherical furnace |
Comparative calibration | -38.8344(mercury point) | 8.0 or less | 500 or more | Fixed point Realizing device |
|
0.01(triple point of water) | 10.0 or less | 400 or more | |||
29.7646(gallium point) | |||||
156.5985(Indium point) | 7.5 or less | 480 or more | |||
231.928(Tin point) | |||||
419.527(Zinc point) | |||||
660.323(Aluminum point) | |||||
961.78(Silver point) | |||||
1084.62(Copper point) | 600 or more |
※ Conditions for shapes, terminals, and terminal accessories are the same as those for comparative calibration.
Please contact us for thermometers of dimensions and shapes other than those listed above.
Calibrated thermometer | Type of calibration | Calibration temperature(℃) | Dimensions of the target thermometer | Equipment used | |
---|---|---|---|---|---|
Outer diameter (mm) |
Length (mm) |
||||
Thermocouple Platinum resistance thermometer Thermometer with indicator |
Comparative calibration | -195.798 | Sheath 8.0 or less Element wire 3.2 or less |
300 or more | Nitrogen boiler |
※ Conditions for shapes, terminals, and terminal accessories are the same as those for the above calibration.
Please contact us for thermometers of dimensions and shapes other than those listed above.
Request for calibration application
・In principle, our thermometer should be brought in to maintain accuracy and prevent damage. In the case of transportation, please be sure to have your advance consultation. However, we cannot transfer the fixed point realizing device and the standard platinum resistance thermometer.
・Depending on the results of the pre-acceptance preliminary test before calibration, calibration as applied may not be possible.
・The period required for calibration varies depending on the condition of the thermometer and the number of temperature points. In case of hurry, please make a prior discussion.