Dimensions of the object that can be calibrated

Calibrated thermometer Type of calibration Calibration temperature(℃) Dimensions of the target thermometer Equipment used
Outer diameter (mm) Length(mm)
Thermocouple
Platinum resistance thermometer
Thermometer with indicator
Comparative calibration -40,0,100,200 Sheath 8.0 or less
Element wire 3.2 or less
300 or more Liquid bath
-196~-80 230 or more 750 or less
Or 850 or more
Drivelock tank
-100~250
(-100~-80)
Sheath 8.0 or less Element wire 3.2 or less
(Sheath 7.0 or less)
350 or more Liquid bath and
Drivelock tank
250~660(arbitrary temperature) Sheath 8.0 or less
Element wire 3.2 or less
400 or more Powder tank
100~1100(arbitrary temperature) 350 or more Tubular furnace
1100~1372(arbitrary temperature) 600 or more
1100~1554(arbitrary temperature) Element wire0.5 or less 550 or more

※ Shape… The length portion of the dimension should have a straight extension shape.
※ Terminal: Have a (lead) terminal that can be engaged and connected with the alligator.
※ Terminal accessories: Accessories such as terminal box and connector can be removed.
Please contact us for thermometers of dimensions and shapes other than those listed above.

 

Calibrated thermometer Type of calibration Calibration temperature(℃) Dimensions of the target thermometer Equipment used
Outer diameter
(mm)
Length
(mm)
Thermocouple
Platinum resistance thermometer
Thermometer with indicator
Comparative calibration 300 ~ 1100(arbitrary temperature) 7.5 or less 200 or more Spherical furnace
Comparative calibration -38.8344(mercury point) 8.0 or less 500 or more Fixed point
Realizing device
0.01(triple point of water) 10.0 or less 400 or more
29.7646(gallium point)
156.5985(Indium point) 7.5 or less 480 or more
231.928(Tin point)
419.527(Zinc point)
660.323(Aluminum point)
961.78(Silver point)
1084.62(Copper point) 600 or more

※ Conditions for shapes, terminals, and terminal accessories are the same as those for comparative calibration.
Please contact us for thermometers of dimensions and shapes other than those listed above.

 

Calibrated thermometer Type of calibration Calibration temperature(℃) Dimensions of the target thermometer Equipment used
Outer diameter
(mm)
Length
(mm)
Thermocouple
Platinum resistance thermometer
Thermometer with indicator
Comparative calibration -195.798 Sheath 8.0 or less
Element wire 3.2 or less
300 or more Nitrogen boiler

※ Conditions for shapes, terminals, and terminal accessories are the same as those for the above calibration.
Please contact us for thermometers of dimensions and shapes other than those listed above.

 

Request for calibration application

・In principle, our thermometer should be brought in to maintain accuracy and prevent damage. In the case of transportation, please be sure to have your advance consultation. However, we cannot transfer the fixed point realizing device and the standard platinum resistance thermometer.

・Depending on the results of the pre-acceptance preliminary test before calibration, calibration as applied may not be possible.

・The period required for calibration varies depending on the condition of the thermometer and the number of temperature points. In case of hurry, please make a prior discussion.

【英語版】山里産業コーポレートサイト
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.