Products

Calibration of practical thermometers

Certification range:-196℃ to 1554℃
Accredited calibrations of industrial thermometers and thermometer calibration devices are available.

Calibrated thermometer Calibration temperature (℃) Expanded uncertainty (℃) Reference standard
Thermocouple
(R, S, B)
-40~660 0.8 Platinum resistance thermometer
300~1100 0.9 R thermocouple
1100~1300 1.6
1300~1554 2.0
Thermocouple
(N, K, E, J, T)
-196~-40 1.2 Platinum resistance thermometer
-40~660 0.6
300~1100 1.0 R thermocouple
1100~1372 1.8
Platinum resistance thermometer
(3-wire, 4-wire, 100Ω)
0
(specified temperature)
0.023 Platinum resistance thermometer
-40, 100, 200
(specified temperature)
0.04
-196~-80 0.14
-80~250 0.10
250~660 0.18
Thermometer with indicator
(Minimum scale: 0.01°C or more)
-195.798 0.12 Platinum resistance thermometer
-196~-80 0.14
-80~250 0.10
250~660 0.18
200~1100 0.6 R thermocouple
1100~1400 1.6
1400~1554 2.6
Thermometer calibration device
(portable type with soaking block, minimum scale 0.1℃ or higher)
-100~250 0.3 Platinum resistance thermometer
250~660 0.6
300~1100 1.2 R thermocouple

※There are restrictions on the shape of the temperature sensor. Check “Dimensions of the object that can be calibrated”.
※Calibration is performed after cleaning the target with ethanol, but depending on the object adhered to the object, the surface may become colored after calibration.
※If preprocessing for calibration is desired, consult with us.
※The maximum calibration capacity varies depending on the condition of the calibrated product (sensor structure, resolution, history, etc.) and calibration conditions (current value, etc.).
※The resistance value of the current 0.5mA is normal for the calibration value of the platinum resistance thermometer. The current can also be specified in 1mA, 2mA.
※The maximum measurement capacity is the value when the thermometer is a platinum resistance thermometer. (thermometer with indicator)
※For the preliminary adjustment of the instrument, a separate technical fee is required (consultation is required). An additional charge will be charged for calibration of both before and after adjustment (thermometer with indicator).
※JCSS standards are guaranteed for traceability, but if you are requested to issue a traceability certificate separately, you will be charged separately.

【英語版】山里産業コーポレートサイト
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.