Products

Radiation thermometer IP140 IPE140

Product name

Digital radiation thermometer

 

Model number

IP140 IPE140

 

IP140/IPE140 is a fast-response (fastest 1.5msec) full-digital radiation thermometer.
Temperature of various metal surfaces and machining processes can be measured with high accuracy by detection at short wavelengths.

Usage

    • Suitable for non-contact temperature measurements such as metal tube type, preheat, annealing, tempering, welding, casting, hardening, sintering, melting, soldering, rolling, brazing, etc.

Features

    • For measuring the temperature of samples with severe temperature changes or moving at high speed because of fast response (fastest 1msec)
      Suitable.
      (However, around the lower temperature limit, the response time is automatically extended to reduce noise.)
    • Four types of movable focal lenses are available to optimize the spot diameter at any distance.
    • A LCD indicator is provided so you can check the temperature indication on the main unit.
      Parameters can be set on the main unit.
      Basic measurements can be made only by supplying power.
    • Small-spot measurement (min. φ0.3mm/70mm) is realized, so even small samples can be measured.
    • The complete digital circuit provides excellent stability and noise immunity.
    • It has a built-in maximum value holding circuit, so it is also suitable for measuring moving objects.
    • The test current output function facilitates connection to an externally installed temperature indicator.
    • Data-recording and parameter setting can be performed using the dedicated measurement software (for Windows) provided as a standard-attached product.

Specifications

The position can be checked with the laser beam or direct-view finder. (Specified when ordering)

Measurement accuracy IP140 2℃ for temperatures below 400℃ and 0.3% + 1℃ of the measured value for temperatures above 400℃
IPE140 2.5℃ for temperatures below 400℃ and 0.4% + 1℃ of the measured value for temperatures above 400℃
Temperature range IP140 50 ~ 400℃, 75 ~ 550℃, 100 ~ 700℃, 160 ~ 1200℃, 200 ~ 1300℃
IPE140 5 ~ 500℃, 30 ~ 1000℃, 50 ~ 1200℃
Detection wavelength IP140 2 ~ 2.8μm
IPE140 3 ~ 5μm

・The interface can also be switched to a RS485 with RS232 and bus-connectivity.

Basic system configuration diagram

5/12/140 Series

Accessories

3834-270 Ball and socket fixing bracket

3834-280-2 L shaped mounting bracket (tripod mountable)

3835-230 Air purge pipe

3837-240 Cooling plate

3837-290-1 Cooling jacket

3835-460 90° mirror (for IPE140)

Scanner for 3843-530 IPE140 (12°, with 5Hz,CaF2 window)
3835-290 Air purge pipe for scanner

3827-200 Emissivity enhancer

3837-140 Ball and socket fixing bracket for cooling jacket

3835-160 Air purge unit for cooling jacket

   

List of Radiation thermometer and thermal image

【英語版】山里産業コーポレートサイト
Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.